GE Healthcare · CTLightSpeed1831083300318328950001
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Scanning aborted with unexpected Loss Of High Voltage.
Error message
Error: 183108 KV fault detected: Cathode shoot-thru. See long description for full error Sun Jun 5 23:17:05 2005 Error: 183108 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 248 Function: XRAY GENERATION : KV Loop Scan: 33003/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 23:17:05:545 Log Series: 424 KV fault detected: Cathode shoot-thru. Fault latch address = ffcff3 hex, Bit = D3 Error: 183289 Unexpected Loss Of High Voltage Wed Aug 4 20:21:15 2004 Error: 183289 Host: OBC Process: Generator File: gen xray.c 1.31 Line: 528 Function: XRAY GENERATION : Rotating HVON Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 20:21:16:325 Log Series: 59 Unexpected Loss Of High Voltage. .
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2218570 BOX OF FUSES for HSA, CTI, and LS 1X - 4X Systems See long description for details 2218570 BOX OF FUSES for HSA, CTI, and LS 1x - 4x Systems Fuse KIT
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.