Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230011073403010440

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230011073 Error detected closing information file. Shutdown processing will continue. Info file: /raw_data/info_file. OS error : 140 --- Remote I/O error

Error: 230011073 STUB -SCANFILE_HEADER_callback See LD for full error Mon Jun 27 09:56:31 2005 Error: 230011073 Host: ct01 Process: headermgr File: sdio_xfs_api.c Line: 4074 STUB - SCANFILE_HEADER_callback Fri Sep 29 13:35:26 2006 Error: 230011073 Host: ct01 Process: badchanmgr File: sdio_xfs_api.c Line: 4101 STUB - SCANFILE_HEADER_callback Error detected closing information file. Shutdown processing will continue. Info file: /raw_data/info_file OS error : 2 --- No such file or directory then repeats ... Fri Sep 29 13:35:26 2006 Error: 230011073 Host: ct01 Process: scanmgr File: sdio_xfs_api.c Line: 4101 Function: Data Acquisition : SBC Processing Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Error detected closing information file. Shutdown processing will continue. Info file: /raw_data/info_file OS error : 11 --- Resource temporarily unavailable then repeats... Fri Sep 29 13:35:46 2006 Error: 230011073 Host: ct01 Process: scanfile_xfer File: sdio_xfs_api.c Line: 4101 Function: Image Generation : SBC Processing Scan 

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.