Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2600074175069014912

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Attention The detector is not at the proper temperature. see long description for more detail. Attention The detector is not at the proper temperature. If you wish to proceed with this scan select the "OK" button, however IQ may be degraded including image artifacts. Please contact GE Service

Error message

Error: 260007417 Detector Temperature Is Low. Image quality may be affected See Long Description for full error Wed Jan 18 07:25:21 2006 Error: 260007417 Host: DAS Process: DCB Firmware File: dtemptst.c Line: 373 Function: DATA ACQUISITION : Detector Heater Scan: 50690/1/2 Type: Axial Exception Level: Sec/Soft Time: 07:25:21:605 Log Series: 51 Detector Temperature Is Low. Image quality may be affected. Current Reading:31.54 Deg C A/D Counts:14912

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2294856-2 (old # 2294856) DHCB assembly Detector Heater Control Board See LD for femc info and picture 2294856-2 (old # 2294856 ) DHCB assembly Comments: FLS - DHCB, BOARD / COVER ASM (H3) DETECTOR HEATER CONTROLLER BD picture Part #: 2294856-2 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: DHCB, Board-Cover Asm EMC ED2 FRU: Yes Status: Orderable Parts Offered as: Current: 2294856-2 Harvested: Not Available Exchange: Not Available Comments: Added per EMC2 RSPL, this part is EMC2 compliant replacement for 2294856. ms13452 050509
Area
CT