Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT200002372202313200002352

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

System software shuts down with various process termination errors.

Error message

Error: 200002372 The process darc_init took too long to start up See Long Description Tue Mar 31 20:23:13 2009 Error: 200002372 Host: hsjk125_ct3 Process: cupMonitor File: ssProcStart.c Line: 856 The process darc_init took too long to start up. Note: You will also see traceback created in the messages log file: cd /var/log more mess* Mar 31 20:23:13 hsjk125_ct3 darc_init: CUP ERROR in cupMakeCore.c line 200 pid 8 118: generating traceback file traceback.darc_init.200931Mar.202313

Error: 200002352 A HOST_SHUTDOWN was initiatied by the unexpected termination of darc_init. See long description for full error Sat Feb 2 07:54:18 2008 Error: 200002352 Host: ct01 Process: cupMonitor File: ssProcStop.c Line: 766 A HOST_SHUTDOWN was initiatied by the unexpected termination of darc_init.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.