GE Healthcare · CTHiSpeed230017082394680230017078503421111522307108580
HiSpeed QX/i - Octane
Symptom
Fastcal is failing and partial ring artifact seen on images. See Long Description for a picture of the artifact Fastcal is failing and partial ring artifact seen on images.
Error message
Error: 230017082 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.3 The row number can be VERY confusing. Please see Long Description for how to decode the ROW NUMBER Tue Mar 22 20:12:04 2005 Error: 230017082 Host: ct01 Process: accs File: aircal_processing.c Line: 917 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.394680 Click Here to Decode the ROW NUMBER Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microseconds = 108580 View scanfileto check data and header information Error: 230023046 Calibration limit check failed. Beam may be obstructed See LD for complete error Tue Mar 22 20:12:04 2005 Error: 230023046 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3953 Calibration limit check failed. Beam may be obstructed
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2336885 MDAS 4 or MDAS 8 Convertor Board, replaces 2322994-3. See Long Description for details. 2336885 MDAS 4 or MDAS 8 Convertor Card, replaces 2322994-3 (see below) Part #: 2336885 Part Type: GP CONSUMABLES Part Weight: 0.48 LBR Repairable: No Returnable: No Hazardous Material: N Description: SERVICE, MDAS CONVERTOR BD FRU: Yes Status: Orderable 4 Slice versus 8 Slice Configuration The HiSpeed QX/i system can operate 4 slice per rotation, data collections mode, whereas, the Detector and MDAS are capable of operating 4 slice or 8 slice per rotation. From a hardware perspective the main difference is in the MDAS. The 4 slice configuration uses a depopulated MDAS. This means that all EVEN slots are empty. Only ODD numbered slots contain converter boards. Data collection is still performed in the same manner using only the odd converter card slot assignments.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required