GE Healthcare · CTDiscovery2601050025013026011823654034260105008

Scanner Hardware Stopped Scan.

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260105002 Startup Vfocus too low. Voltage has been turned off. See long description Tue Mar 16 06:56:56 2010 Error: 260105002 Host: orp Process: orp_subsys File: gen_exec.cxx Line: 2792 Tue Mar 16 06:56:55 2010 Host : Jedi Ermes # : 260105002 Exception Class : Abort Severity : Primary File : gen_exec.cxx Line# : 2792 Function : No System Function Reported Scan Type : Axial Scan: 50130/1/2 (50-0242) Startup Vfocus too low.Voltage has been turned off. Focal spot width affected. Z6,Z7= 0h,0h

Error: 260118236 - (70-0501) DC bus out of range Wed Nov 26 12:56:16 2003 Error: 260118236 Host: orp Process: orp_subsys File: gen_exec.c Line: 1923 Wed Nov 26 12:56:14 2003 Host : Jedi Ermes # : 260118236 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 1923 Function : No System Function Reported Scan Type : Static Scan: 54034/1/4 (70-0501)DC bus out of range. Z6,Z7= 0h,0h

Error: 260105008 (20-0274) Vfocus too low. Focal spot width affected. See long description Sun Sep 6 17:16:26 2009 Error: 260105008 Host: orp Process: orp_subsys File: gen_exec.cxx Line: 2811 Sun Sep 6 17:16:24 2009 Host : Jedi Ermes # : 260105008 Exception Class : Secondary Severity : Pri/Soft File

Applies to

Models
Discovery
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5332530 SC-Fast KV HV Tank for Discovery HD- Use after option upgrade ONLY See long description for details Part #: 5332530 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: SC-Fast KV HV Tank FRU: Yes Status: Orderable Comments: ** FE ASSIST ** Be sure you have SW version 09MW29.14. If you have SW version 08MW39.4 then you must order 5173970-2 SC HIGH VOLTAGE TANK. (212001147 010510) *** 2ND OPINION NOTES *** As of 02/11/10 Call Engineering Contact John Breunissen - 414-202-9485 or Daniel Nunez - 414-721-8460 Before ordering. Also, perform: Meter verification, No load test with/without tube, Focal spot diagnostics(with/without kV), and Filament heating diagnostic. If any of these tests fail without tube connected, replace the tank and redo the tests. Per HDVCT Options RSPL, 212008898, 5/4/09. 2ND OPINION NOTES *** Before ordering HV Tank perform: Meter verification, No load test with/without tube, Focal spot diagnostics(with/without kV), and Filament heating diagnostic. If any of these tests fail without tube connected, replace the tank and redo the tests.
Area
CT

Full cause & step-by-step resolution