Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26012101326694

LightSpeed RT16 & RT16 Xtra and Vision

Symptom

Gantry tilt moves a small distance then stops (less than 2 deg) Gantry tilts forward a small distance then stops

Error message

Error: 260121013 Gantry velocity less than the minimum allowed velocity. See Long Description for more details Wed Mar 29 11:28:24 2006 Error: 260121013 Host : Table Gantry Processor Ermes # : 260121013 Exception Class : Secondary Severity : Pri/Soft File : tilt mgr.c Line# : 1171 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Gantry velocity less than the minimum allowed velocity. Gantry velocity:-2 mdeg/sec Tilt position:-26694 mdeg Minimum allowable velocity:300 mdeg/sec

Applies to

Models
LightSpeed, VCT
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues