Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2600010223494926000107150039

LightSpeed 4.X - GRE

Symptom

Scan aborts WITH Communications with the CCB sub-system has not been established. Scan aborts WITH Communications with the CCB sub-system has not been established

Error message

Error: 260001022 Communications between the CCB sub-system and OBC have been lost. Attempting to establish communications See long description for full error Fri Feb 10 14:15:14 2006 Error: 260001022 Host: OBC Process: HOP File: hop_init_state.c Line: 249 Function: No System Function Reported Scan: 34949/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 14:15:13:720 Log Series: 600 Communications between the CCB sub-system and OBC have been lost. Attempting to establish communications.

Error: 260001071 The CCB has reset unexpectedly. The OBC has successfully reconnected to sub-system. See LD for complete error Error: 260001071 The CCB has reset unexpectedly. The OBC has successfully reconnected to sub-system. Fri Feb 10 16:23:44 2006 Error: 260001071 Host: OBC Process: HOP File: hop_node_init.c Line: 347 Function: No System Function Reported Scan: 50039/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 16:23:44:745 Log Series: 130 The CCB has reset unexpectedly. The OBC has successfully reconnected to sub -system. Reason for reset as reported by CPU: External signal source. CPU reset register: 0x80 Node status: 0x382 State flags: 0x9800682

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.