Symptom

Hardware reset fails with Laser indicator flashing on the Gantry Display board. Hardware reset fails with Laser indicator flashing on the Gantry Display board

Error message

Error: 244634 Subsystem State; DCB:down, CCB:down, TCB:down, JEDI:down, TGP/ORP:down, DEVICES:up. GOC6 version All Down except Devices See Long Description Mon Dec 8 14:32:41 2008 Error: 244634 Host: CTUH07 Process: fwmgr File: fwmgr_utils.c Line: 677 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:down CCB:down TCB:down JEDI:down TGP/ORP:down DEVICES:up . The error above is from a GOC6, devices up is new and this stands for new feature like intergrated injector which has not been release at this time ( 12/23/2008)

Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state See Long Description for complete error Sun Sep 26 08:36:46 2004 Error: 244540 4 000ctge04 fwmgr FwMgrEventHandler.c 1517 Running diagnostic: Scan Data Path Functional Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5115622-2 replaces 5115622 ORPV BOARD ASSY See Long Description for more information (Part number changed after FMI 25392 to 5115622-2) 5115622 ORPV BOARD ASSY pictures below Part #: 5115622 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: ORPV BOARD ASSY Positioning LS32 FRU: Yes Status: Orderable Comments: Added per VCT RSPL, ms13452 041209 Set as Critical Part for Product CT_VCT, ms04278, 05Oct14
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.