GE Healthcare · CTLightSpeedVCT290010005290009007290009011
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Prospective images slow to reconstruct. Some images came up and then a minute later the other images showed up.
Error message
Error: 290010005 Socket error: socket is terminated because IG dies (error text varies) See Long Description for error examples. Tue Mar 16 10:40:09 2004 290010005 darc data_restore request_object.cxx 503 Socket error: socket is terminated because IG dies Fri Nov 16 19:22:57 2007 Error: 290010005 Host: darc Process: data_restore File: request_object.cxx Line: 556 Socket error: socket event from IG Error: 290009007 Timeout waiting for image available. Number of retry so far: 1. IG node: 0. Sun May 14 14:05:42 2006 Error: 290009007 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 5208 Function: Image Generation : SBC Processing Job Type: RECON_JOB_PROS_HELICAL_RECON Scan: 378/2/2 Image: 378/2/39 Timeout waiting for image available. Number of retry so far: 1. IG node: 0. Error: 290009011 Timeout waiting for abort response from the IG. Will reset the IG (1,2,3) See Long Description for more details Error: 290009011 Timeout waiting for abort response from the IG. Will reset the IG. (1,2,3) Tue Aug 15 13:00:05 2006 Error: 290009011 Host: darc Process: recon_control File: recon_control_cb.cxx Line: 1066 Timeout waiting for abort response from the IG. Will reset the IG.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5308760 Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible - used On GOC6 Disk Array See Long description: Part #: 5308760 Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible used on GOC6 Disk array ( 5308759 ). Part #: 5308760 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible with 5308759 FRU: Yes Status: Orderable Comments: Per HDVCT RSPL, 212008898, 7/15/08. ** FE ASSIST ** This is a single Hard Drive used in the High Speed Disk Array on the GOC6 Console CT 750 HD and VCT (212001261 23-Jan-2009) ST32050310NS = 250 GB SATA II Hard Disk Drive
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.