Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26000681723391

LightSpeed 4.X - Octane

Symptom

Scanning aborts with axial encoder errors

Error message

Error: 260006817 Axial Motor Control (AMC) Logic detected Gantry (Axial) MOTION without Command Current Position: 108.7 degs Commanded Position: 90.0 degs Degrees Tol: +/- 1 See the long description for the full error code Mon Dec 6 09:27:53 2004 Error: 260006817 Host: STC Process: Axial File: amc_mon.c Line: 987 Function: AXIAL CONTROL : axial control Scan: 23391/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 09:27:53:795 Log Series: 829 Axial Motor Control (AMC) Logic detected Gantry (Axial) MOTION without Command Current Position: 108.7 degs Commanded Position: 90.0 degs Degrees Tol: +/- 1 Retry Rx to verify probable Gantry Balance or Brake or Encoder Hardware. problem.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues