GE Healthcare · CTLightSpeedVCT2601250418489528457501241904294907296445950
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260125041 Uncommanded Table Motion with the Elevation drive. See long description for a sample error message. Thu Jan 28 08:29:19 2010 Error: 260125041 Host: tgp Process: tgp subsys File: ta callback gt.c Line: 504 Thu Jan 28 08:29:19 2010 Host : Table Gantry Processor Ermes # : 260125041 Exception Class : Abort Severity : Primary File : ta callback gt.c Line# : 504 Function : No System Function Reported Scan Type : Axial Scan: 1212/2/9 Uncommanded Table Motion with the Elevation drive. Current pos:848952 um Last stop pos:845750 um Tolerance:3000 um Time since last stop:124190 ms Last target pos:4294907296 um Hardware status:0 Another drive (Elevation or Cradle) pos:445950 um
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5127774 ELEVATION POTENTIOMETER ASSY-VCT See Long Description for details Part #: 5127774 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION POTENTIOMETER ASSY FRU: Yes Status: Orderable Comments: source pole/planner code revised per to Global Table transfer
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.