GE Healthcare · CTLightSpeedVCT1831772300000675000416800015000230000065
LightSpeed 1.X - GRE
Symptom
Scanning aborted with Buffer overrun detected during scan data transfer
Error message
Error: 183177 Abnormal Xray termination. External x-ray interlock opened See the long description for the full error code Tue Sep 6 09:47:22 2005 Error: 183177 Host: OBC Process: Generator File: gen_xray.c 1.31 Line: 604 Function: XRAY GENERATION : Xray Exposure Interlocks Scan: 637/7/1 Type: Helical Exception Level: Pri/Most Severe Time: 09:47:22:260 Log Series: 487 Abnormal Xray termination. External x-ray interlock opened. Error: 230000067 Unexpected event occurred during scan data transfer See the long description for the full error code Sat Apr 17 20:34:13 2004 Error: 230000067 Host: darc Process: dataacq File: dip_thread.cxx Line: 390 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Xray-on Static Scan: 50004/4/1 Scan Seq Id: 168000 Unexpected event occurred during scan data transfer. Event bit: 7 Count: 1 Description: Global bus error: read attempted from the Test Data FIFO Scan aborted. OR Tue Mar 6 07:13:50 2007 Error: 230000067 Host: darc Process: dataacq File: vdip.cxx Line: 817 Function: Data Acquisition : Scan Data Acquisition Scan Type: Cine Scan: 50004/6/1 Scan Seq Id: 15000 Unexpected event occurred during
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2117848-2 24 meter DAS Fiber Optic Cable (long) used on HSA through Lightspeed 1.x thru 5.x See LD for FEMC information P/N 2117848-2 24 meter DAS Fiber Optic Cable (long) used on HSA through Lightspeed 1.x thru 5.x from the Gantry to the OC Console. Part #: 2117848-2 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: 24384 +/- 150 FIBER OPTIC CABLE FRU: Yes Status: Orderable Parts Offered as: Current: 2117848-2 Harvested: 2117848-2U Exchange: Not Available Comments: FLS - FIBER OPTIC CABLE USED ON HSA RP AND CT/I 1 USE IS FROM SCOM-STC IN GANTRY TO FEP IN CONSOLE LENGTH 24384 MM GELS - 03/09/09 IF NO STOCK PLEASE USE 2117848-2U BUT NO PDR BOURGFL 973,FIBER CABLE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required