Pular para o conteúdo principal
GE Healthcare · CTHiSpeedLightSpeed200002730527288200000411

Scout images fail to reconstruct.

Symptom

Scout images fail to reconstruct. See long description for details Scout images fail to reconstruct. Rebooting, cycling console power, etc. does not help.

Error message

Error: 200002730 Image space allocate response size error. See long description for details Sun Jun 3 10:48:22 2007 Error: 200002730 Host: ct01 Process: imagecreate File: ic_imagecreate_req.c Line: 1284 Image : Image space allocate response size error Requested size : 527288 Response path count : 1 Response status : -1

Error: 200000411 Image space allocation failed Tue Apr 10 06:22:01 2007 Error: 200000411 Host: arls Process: dms AllocSpace.cxx Line: 212 Image space allocation failed.

Applies to

Models
HiSpeed, LightSpeed
Product
HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) See FMI 25903 I Helios_H2_O2LESS_CLASSC_Product 09/21/2005 Helios ver 1104L.3 rel 1104L.3_H4.2CLASSC I Helios_CLASSMOC_Product 09/21/2005 CLASSMROOT ver 1204L.3 rel 1204L.3_H4.2CLASSM Class C is 2394897 Order the following P/N: 2401759 (6 pack) Class M is 2394898 Order the following P/N: 2401759 (6 pack) Software Collector 5115431 Recon Patch for GOC2 only 5123440
Area
CT