Pular para o conteúdo principal
GE Healthcare · CTHiSpeed24506911185760011458314577

Scanner Hardware Stopped Scan.

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 245069 Aborting xrays during helical views expected xxxxx - views collected yyyyy See long description Thu Dec 16 14:03:04 2010 Error: 245069 Suite: UVC2 Host: SBC0 Process: scanmgr File: do_continuous_complete.m 1.44 Method: do_continuous_complete() Line: 391 Function: Data Acquisition : SBC Processing Scan: 11185/2/1 Type: Helical Scan Seq: 76001 Aborting xrays during helical because firmware reports no scanning error, but dataacq did, views expected 14583 - views collected 14577

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2302509-2 SIGNAL BRUSH BLOCK FOR HSA/CTI SYSTEMS See long description for details Part #: 2302509-2 Part Type: GP CONSUMABLES Part Weight: 0.62 LBR Repairable: No Returnable: No Hazardous Material: N Description: SIGNAL BRUSHBLOCK - REPLACEMENT CT/I FRU: Yes Status: Orderable Comments: FLS - 8-7-03, CONSUMABLE COLLECTION PER ANJANA ON PROMOTION THROUGH JUNE 30, 2004 DO NOT CHANGE PRICE PER SHAWN (MAYND) DO NOT DISCOUNT - PER SHAKUN Note: 2302509 FOR H1 (LIGHTSPEED) SYSTEMS That is, DO NOT mix and match ETC and Schliefring Signal Brush Blocks. Any matched pair of Signal Brush Blocks may be installed with either Power Brush Block. See HSA/CTI Non EMC and EMC Gantry Slipping Comparison in Direction 46-018316 for any updates to this information.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required