Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18418250076260007323

LightSpeed 4.X - GRE

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 184182 DIP Ready-For-Views message received before offset triggers were stopped. See long description for a full sample error message. Wed Apr 22 08:20:42 2009 Error: 184182 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1648 Function: No System Function Reported Scan: 50076/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 08:20:42:925 Log Series: 45 DIP Ready-For-Views message received before offset triggers were stopped. HW/SW Error: Probable problem with das trigger or das view hardware or cables. Possible sofware sequencing problem.

Error: 260007323 Unacceptable Amount of Acquisition Trigger Jitter See LD for complete error Fri May 14 07:25:51 2004 Error: 260007323 Host: DAS Process: DCB Firmware File: abort.c Line: 315 Function: DATA ACQUISITION : Scan Data Trigger Generation Scan: 4371/1/1 Type: Axial Exception Level: Sec/Soft Time: 07:25:51:075 Log Series: 51 Scan ABORT -- DAS Has Detected An Error.^B Acquisition hardware (DCB) has detected an Unacceptable Amount of Acquisition Trigger Jitter. (Trigger Frequency: 984 hertz). (State: Normal State Command Level: Scan in Progress ScanState: Collecting Offsets State)

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues