Pular para o conteúdo principal
GE Healthcare · CTBrightSpeedVCT200109130260111601

BrightSpeed HPower, Delight, Velocity or MERI HPower

Symptom

Step and Shoot (SmartStep or SmartView) hand controller not functioning See LD for more Step and Shoot (SmartStep or SmartView) hand controller not functioning When customer attempts to use the hand controller for VCT step and Shoot, the system will not initialize into step and shoot mode.

Error message

Error: 200109130 Received a bad reply status from Firmware for Enabling The IHHC See long description for a full sample error. Wed Jul 2 11:44:39 2008 Error: 200109130 Host: ct06 Process: scanRx File: uifIpcMgrScanRx.c Line: 7673 Function: Data Acquisition : OC Processing Received a bad reply status from Firmware for Enabling The IHHC, status = -1, se q num = 636. Please Check IHHC connections.

Error: 260111601 IHHC reset timeout. See LD for complete error Thu Oct 20 12:41:32 2005 Host : Table Gantry Processor Ermes # : 260111601 Exception Class : Abort Severity : Primary File : hc_exec.c Line# : 766 Function : No System Function Reported Scan Type : None/Unknown Scan: 4628/4/0 IHHC reset timeout.

Applies to

Models
BrightSpeed, VCT
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2199947 CT SONATA HAND HELD CONTROLLER CT SONATA HAND HELD CONTROLLER Part number 2199947 CT SONATA HAND HELD CONTROLLER Part number 2199947
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.