GE Healthcare · CTLightSpeed18220870636654301049466531
LightSpeed 4.X - GRE
Symptom
Cradle will not drive in or out. Cradle will not drive in or out
Error message
Error: 182208 Alignment error between Encoder and Potentiometer. Potentioment Position 70636 (.01 mm) Encoder Position 65430 (.01 mm) Tolerance 2540 (.01 mm) See the LD for the complete error code Error: 182208 Alignment error between Encoder and Potentiometer. Tue Aug 5 15:05:36 1997 Suite: PCT2 Host: ETC Proc:Cradle Error: 182208 File: cmcEncoderRef.c 1.8 8/27/ Method: No Method Line: 152 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 10494/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:05:35:935 Log Series: 27 HW error: Encoder C Pulse Alignment error with current Potentiometer reading. Current Position 66531 (.01 mm). Encoder Alignment Minimum (1456) and Maximum (1658) Potientiometer Window based on Last Potentiometer Reading (1658).
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-297036G2 longitudinal potentiometer See LD for more info 46-297036G2 longitudinal potentiometer ( NOTE the pot/encoder assy is # 46-296234G1 ) Part #: 46-297036G2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: POT ASSY. FRU: Yes Status: Orderable Comments: FLS - PF8 TWICE FOR FE ASSIST COMMENTS...... COLLECTION OF DEFECTIVES START7/23/01 MS5277 SPP CODE CHANGED FROM 2 TO N PER PCN 220431. MSA8429 03/25/02 PF8 AGAIN FOR FA ASSIST COMMENTS. FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE "U" PART IF AVAILABLE. IF NO "U" USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART. FLS - FE ASSIST (971512) (MS6588) 46-297036G2 IS THE HSA POTENTIOMETER ASSEMBLY WHICH CONSISTS OF A BURNS 5K POT AND A 2.2 MICRO FARAD CAP. THE CAPACITOR IS SOLDER ACROSS THE WIPER AND RETURN TERMINALS. THE 9800 TABLE ELEVATION POT (46-165135P3) IS THE SAME 5K BURNS POT BUT DOES NOT HAVETHE NOISE REDUCING CAPACITOR PLEASE REFER TO SN-T1336 FOR DETAILS REGARDING CAPACITOR AND POT NOISE DO YOU STILL WISH TO ORDER THIS PART?
- Area
- CT