GE Healthcare · CTLightSpeedVCT260134729260134714260134715834158
LightSpeed RT16 & RT16 Xtra and Vision
Symptom
Table will not move down; with tape or touch switch error See Long Description for switch locations. Table will not move down; tape switch was either contacted or not present The VCT table has a large number of touch sensors that can produce a variety of error codes.
Error message
Error: 260134729 Elevation cannot move because any of the touch switches are already contacted or disconnected. See long description for full error Error: 260134729 Host: tgp Process: tgp_subsys File: gtcb_emc_profile.cxx Line: 166 Tue Jun 20 14:59:35 2006 Host : Table Sub-System Ermes # : 260134729 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_profile.cxx Line# : 166 Function : No System Function Reported Scan Type : None/Unknown Scan: 3261/4/0 Type of the touch switches will be shown above log message. - Error: 260134714 Error: 260134715 Elevation touch switch3 (switch on underside of top rear cover) is contacted. See Long Description for complete error Mon Nov 21 14:14:07 2005 Error: 260134715 Host: tgp Process: tgp_subsys File: gtcb_emc_hardware.cxx Line: 714 Mon Nov 21 14:14:07 2005 Host : Table Sub-System Ermes # : 260134715 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_hardware.cxx Line# : 714 Function : No System Function Reported Scan Type : None/Unknown Scan: 367/2/0 Elevation touch switch3 (switch on underside of top rear cover) is contacted. Position is 834158 um from the table base.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution