GE Healthcare · CT26000271050407
Scanner Hardware Stopped Scan
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260002710 DAS Converter Board(s) IIC Communication READ Error. Sat Apr 16 13:28:48 2005 Error: 260002710 Host: DAS Process: DCB Firmware File: cnvread.c Line: 198 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 50407/1/3 Type: Axial Exception Level: Pri/Soft Time: 13:28:48:630 Log Series: 44 DAS Converter Board(s) IIC Communication READ Error.^C Converter Board Bit Map:0x000000000080. Bit Map Indicates Failed Board(s) for Multiple Board Errors. (Board Bit Map defines: BIT47 = CnvBrd 48 ... BIT0 = CnvBrd 1) Bus1: Brd:0 Stat:0 Dev:0 Addr:0x0. Bus2: Brd:8 Stat:102 Dev:2 Addr:0x0. (Device - 1:EEPROM 2:Temp 3:Micro)
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2137958-4 SDAS Convertor Board (replaces 2137958-3) May use 2137958-3U converter Board if no current stock of the 2137958-4 See long description for details Part #: 2137958-4 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: -39999.60 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: CONVERTOR WITH BLACK BEUTY CONNECTOR FRU: Yes Status: Orderable Comments: OBSOLETE PART case 10788898 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support 212012173 06-26-07 REPLACES P/N 2137958-3 THE NEW -4 IS COMPATIBLE TO THE -3. MS20544 ON 3/7/03. ** FE ASSIST ** May use the USED part 2137958-3U converter Board if no current stock of the 2137958-4. (9/10/2007 212001261)
- Area
- CT
Full cause & step-by-step resolution