Pular para o conteúdo principal
GE Healthcare · CT23000350940115230004022

Lightspeed 1.X - Octane

Symptom

Images would not reconstruct.

Error message

Error: 230003509 Ecomm error: Timed out while attempting to receive a message See long description for error variations Thu Sep 11 10:27:19 2008 Error: 230003509 Host: ct02_sbc Process: igctrl File: igproxy_if.c Line: 659 Function: Image Generation : SBC Processing Job Type: Prospective Helical Scan: 40115/2/1 Image: 40115/3/33 Recon Seq Id: 762 ecomm_send_and_wait_for failed sending 9208 to igproxy and waiting for 9209 Ecomm error: Timed out while attempting to receive a message. IG unit=0 resetting IG Thu Sep 11 10:27:24 2008 Error: 230003509 Host: ct02_sbc Process: igctrl File: igproxy_if.c Line: 244 Function: Image Generation : SBC Processing Job Type: Prospective Helical Scan: 40115/2/1 Image: 40115/3/33 Recon Seq Id: 762 ecomm_send_and_wait_for failed sending 9202 to igproxy and waiting for 9203 Ecomm error: Timed out while attempting to receive a message. IG unit=0 closing IG

Error: 230004022 IG Self Test Failure DSP0 with error IG See long description for details Error: 230004022 IG Self Test Failure DSP0 with error IG Mon Jul 27 18:35:30 2009 Error: 230004022 Host: ice Process: igproxy File: igproxy events.c Line: 709 IG Self Test Failure DSP0 with error IG cpu self test 

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2118899-2 RHAPSODE IMAGE GENERATOR Board (IG Board). Used in Lightspeed 1x, CTi 4x, CTi 6x systems. See LD for details Part #: 2118899-2 Part Type: GP REPAIRABLES Part Weight: 3.64 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: RHAPSODE IMAGE GENERATOR FRU: Yes Status: Orderable Comments: FLS - IMAGE GENERATOR BOARD USED ON RHAPSODE MS1356 BURR 12-95 REPLACED 2118899 PER PCN 189598. MS4636 2/6/96 PCN 189939 GELS - KANBAN CSS F00,PCB IMAGE GENERATOR
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.