Symptom

Collimator Cal fails in fastcals. Collimator Cal fails in fastcals

Error message

Error: 230017813 ACCS CCal: Warning: Z target set manually with Thin Twin correction See long description for sample error. Sat Sep 11 17:02:20 2010 Error: 230017813 Host: yslct16 Process: accs File: collSweepProcessing.c Line: 845 ACCS CCal: Warning: Z target set manually with Thin Twin correction

Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1

Error: 230017823 Error: Invalid ratio range detected during the Fast Cal See the long description for the full error Thu Aug 26 02:56:40 2004 Error: 230017823 Host: ct01 Process: accs File: collSweepProcessing.c Line: 2436 ACCS CCal: Error: Invalid ratio range detected during the Fast Cal. The minimum ratio is 1.038304 and is great than the maximum ratio 0.900000

Error: 230023045 The Collimator calibration drift check has failed See the long description of the full error code Thu Aug 26 02:56:40 2004 Error: 230023045 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3155 The Collimator calibration drift check has fai

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) ConsoleClass C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2394897) 1104L.3_ H4.2CLASSC LightSpeed/ HiSpeed QXi Class C Software M4 5111370 5111371 LightSpeed 4.X Applications Software 405L.8_H4.2M4 FW11-04 FMI 25338
Area
CT

Full cause & step-by-step resolution