GE Healthcare · CTHiSpeed150004579918200001340
Image reconstruction keeps failing.
Symptom
Image reconstruction keeps failing. Eventually the recon status goes into shutdown. See long description for details Image reconstruction keeps failing. Eventually the recon status goes into shutdown. Retro recons also fails to recon. When attempting to go into recon management a pop up error message comes up that says “error in reading queue”. Rebooting the system software fails to clear up the problem.
Error message
Error: 150004 Can't find entry xxxxxx in queue /usr/g/db/queues/reconQ See long description for complete error Tue Feb 22 08:10:00 2005 Error: 150004 Suite: MSE1 Host: SBC0 Process: qmgrMain File: QueOnDisk.m 1.67 Method: nextInactive: Line: 1439 Can't find entry 579918 in queue /usr/g/db/queues/reconQ Error: 200001340 Failed to get a synopsis of a queue id See the LD for the complete error code Error: 200001340 Failed to get a synopsis of a queue id Thu Dec 16 15:07:24 2004 Error: 200001340 Suite: CT01 Host: SBC0 Process: scif File: scif_queue_synopsis.m 1.7 Method: get_queue_snapshot_and_synopsis Line: 219 Function: System Control : SBC Processing Failed to get a synopsis of a queue id.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 4.4.19.1 for HiSpeed CT/i with with CT/I (RHAPSODE) CONSOLE (Indigo) - showprods returns Rhapsode Product 4.4.19.1 4.4.19.1 for HiSpeed CT/i with with CT/I (RHAPSODE) CONSOLE (Indigo) - showprods returns Rhapsode Product 4.4.19.1 Intro. Date 10/18/00 Mandatory FMI 25268 Rhapsode Product 4.4.19.1 software for HiSpeed CT/i with SGI Indigo computer Intro. Date 10/18/00 Mandatory FMI 25268 Application Software Set Set Part No. . 2283723-3 Apps. CD Part No 2282440-3 Apps. Ver. 4.4.19M4 Apps. CD Label HiSpeed CT/I App SW showprods/Service Desktop Rhapsode Product 4.4.19.1 Class C CD Part No. Ver. CD Label 2282441-2 4.419M4 HiSpeed CT/I Class C Software Class M CD Part No. Ver. CD Label 2282442-2 4.419M4 HiSpeed CT/I Class M Software Patch Info: P/N 2318818, 4.5 Patch Rhapsode Product 4.4.19.1 Rhapsode patch 4.5.3 ver 1 11/21/2001 Optional FMI 25287A Also available via AutoSC ,_,Number:,2283723-3_______, ,BOM,Rev:,___,PI Rev:,___,CR Nb/Typ:,_______,___, ,CTZ,US, ,JAROSCH,ANN,K , ,Rev:,000,PI Rev:,000,CR Nb/Typ:, , , ,Description/Name:,E,HISPEED CT/I OPER SFWR VER 4.4.19 M4, ,Date:,2000-11-13, ,AIL:, , ,R,S,S,F, ,0001,/,0002, ,Mark, ,Number, ,Quantity,,U/M,U,T,P,C,Description/Name, ,__,1 ,2282439-2 , 1.000,PC ,S,N,N,N,IRIX 6.2M4 OS FOR 4.4 SOF, ,__,2 ,2282440-3 , 1.000,PC ,S,N,1,1,HISPEED CT/I APP SFWR VER, , , , , , , , , , , ,
- Area
- CT
Full cause & step-by-step resolution