Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260007391

"Scan hardware Stopped Scan" with DAS power errors.

LightSpeed 4.X - Octane

Symptom

"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors

Error message

Error: 260007391The Data Acquisition System Has Detected An Error In The Analog Ground 2 DAS Power Supply. Wed Apr 5 10:58:29 2006 Error: 260007391 Host: orp Process: orp_subsys File: pwrtst.c Line: 90 Wed Apr 5 10:58:28 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The Data Acquisition System Has Detected An Error In The Analog Ground 2 DAS Power Supply. Measured: 686 mV, Expected range: -100 to 100 mV

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2334456 12 volt power supply Asm - used on MDAS16 See Long Description Part #: 2334456 Part Type: GP CONSUMABLES Part Weight: 6.17 LBR Repairable: No Returnable: No Hazardous Material: N Description: PWR SPLY ASM, 12V DC (H16) FRU: Yes Status: Orderable Comments: FLS - PWR SPLY ASM, 12V DC (H16). ADDED PER H16 PROGRAM BY MS20544 ON 7/25/02.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues