Pular para o conteúdo principal
GE Healthcare · CTLightSpeedBrightSpeedHiSpeed26011821150018

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118211 ( 60-0313 )Inverter max. ILR current detected. This is a hardware detection of maximum current in serial resonant circuit. Z6,Z7= 0h,0h See long description for complete error Mon Feb 13 17:22:21 2006 Error: 260118211 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Mon Feb 13 17:22:20 2006 Host : Jedi Ermes # : 260118211 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50018/1/4 (60-0313)Inverter max. ILR current detected. This is a hardware detection of maximum current in serial resonant circuit.

Applies to

Models
LightSpeed, BrightSpeed, HiSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2401826-2 Jedi 60 Shielded Power Unit / High Voltage Tank Assembly FRU Package for G-DAS systems Includes: 2377668-2 Jedi 60 Power Unit 2186730 Jedi 60 HV Tank Shielding improved for G-DAS systems See LD for details 2401826-2 JEDI 60DC SHIELDED POWER UNIT / HV TANK HLB ASSY With FRU PACKING FOR G-DAS SYSTEMS (Backward compatible with M-DAS systems) Part #: 2401826-2 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI 60DC SHIELDED POWER UNIT / HV TANK HLB ASSY With FRU PACKING FRU: Yes Status: Orderable Comments: Revised per to MFG confirmation. This is part of Jedi Collector # 2371333-2 INCLUDES: 2377668-2 JEDI 60DC HLB POWER UNIT with INVERTER and PPC KV Control 2186730 JEDI 60 STD HV TANK
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues