GE Healthcare · CTLightSpeedVCT26010720426010720650180

LightSpeed 5.X (60kW)

Symptom

"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors

Error message

Error: 260107204 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry. See Long Description for full error Wed Apr 7 21:18:13 2004 Error: 260107204 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 230 Wed Apr 7 21:18:13 2004 Host : Table Gantry Processor Ermes # : 260107204 Exception Class : Abort Severity : Secondary File : amc_mgr.c Line# : 230 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry.

Error: 260107206 Axial Motor Control (AMC) logic will NOT accept commands until the Axial Drive is properly configured/initialized. See Long Description for full error Wed Apr 7 21:04:02 2004 Error: 260107206 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3771 Wed Apr 7 21:03:47 2004 Host : Table Gantry Processor Ermes # : 260107206 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3771 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic will NOT accept

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
Area
CT