GE Healthcare · CT
ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt)
Symptom
Scan aborts with Aperture error.
Error message
ER-100 Aperture error see long description ER-100 aperture error ER–100 X Aperture ErrorError Code X =1: Aperture Request Command Error Received an undefined communication command. Check the firmware revisions on the DMP4,OC INTF3, TGP, and OGP boards.2: Aperture 1 mm Set ErrorCheck that the position of the optical switch is correct, that the spring to retract the apertureworks normally, and that nothing hinders the moving mechanism. Also check the cable connectors for the sensors and the motor.3: Aperture 2 mm Set ErrorCheck the items described above.4: Aperture 5 mm Set ErrorCheck the items described above.5: Aperture 10 mm Set Error Check the items described above.6: Aperture 3 mm Set ErrorCheck the items described above
Applies to
- Product
- ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2172789-2 P9204LB APERTURE ASSY see long description 2172789-2 P9204LB APERTURE ASSY Part #: 2172789-2 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: P9204LB APERTURE ASSY FRU: Yes Status: Orderable Parts Offered as: Current: 2172789-2 Harvested: Not Available Exchange: Not Available Comments: added per to ECR2091969 by sso#305010201 in Sep 25, 2009
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required