Symptom

Scanner failed to scan with Image Allocation errors See Long Description for error examples Scanner failed to scan with Error: 2021 mkdir failed /usr/g/sdc_image_pool Input/output error and Error: 210000434 IMAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data.

Error message

Error: 200109100 UIRx_xxxxxxxxxx: a failure had been detected while calling UIRx_xxxxxxxxxx function. See LD for error examples. Tue Oct 19 09:22:20 2004 Error: 200109100 Host: dgct Process: scanRx File: UIRxToScanner.cxx Line: 5124 Function: Data Acquisition : OC Processing UIRx_buildGroup: a failure had been detected while calling UIRx_allocateIma geSpace function. Tue Oct 19 09:22:20 2004 Error: 200109100 Host: dgct Process: scanRx File: UIRxToScanner.cxx Line: 5435 Function: Data Acquisition : OC Processing UIRx_ToScanRequest: a failure had been detected while calling UIRx_buildGro up function.

Error: 200502003, Cannot allocate image space for XXX images, allocated XX images. See LD for complete error Mon Feb 13 16:36:28 2006 Error: 200502003 Host: ls16dn Process: retroRecon File: srx_sma_manage_image_space.c Line: 690 Function: No system function reported Cannot allocate image space for 148 images, allocated 7 images.

Error: 210000441 Image space allocation failed. See LD for complete error Mon Mar 26 12:51:14 2007 Error: 210000441 Host: roct Process: dms File: Initialize.cxx Line: 856 Image space allocation failed. size 3863552 Error code: -1

Error: 210000434 IMAGE ALLOCAT

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.