LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Fastcal is failing with beam obstructed errors. Fastcal is failing with beam obstructed errors

Error message

Error: 230023046 Calibration limit check failed. Beam may be obstructed See LD for complete error Tue Mar 22 20:12:04 2005 Error: 230023046 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3953 Calibration limit check failed. Beam may be obstructed

Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microseconds = 108580 View scanfileto check data and header information

Error: 230017083 Possible beam obstructed - sin/cos vector magnitude = 0.xxxxxx - may include row and channel numbers See LD for complete error Mon Oct 31 07:08:03 2005 Error: 230017083 Host: cpcvct2p Process: accs File: aircal_spectral_acal_util.c Line: 445 Possible beam obstructed - sin/cos vector magnitude = 0.385870 Mon Mar 21 07:27:12 2011 Error: 230017083 Host: sj01_ct64s_6001 Process: accs File: aircal_spectral_acal_util.c Line: 449 Possible beam obstructed - row 58(26B), detector channel 375, sin/cos vector magnitude = 0.381015

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2203602 (replaces 2161307) HEMIT HV TANK used on CT/I (performix), HiSpeed QXi, LS1.X LS2.X, LS 3.X, LS4.X See Long Description for details Part #: 2203602 Part Type: GP REPAIRABLES Part Weight: 55.12 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: HEMIT HV TANK SPARE PART FRU FRU: Yes Status: Orderable Comments: ** FE ASSIST ** The HV adapter/Spacer part number 2171866 Kit that is included with the HV tanks But it is only used on older Highspeed Gantry due to the smaller bolt size. Refer to MI&CT ESERVICE NOTE SN07091114a for additional details. (212001261 21-Jan-2009) *** 2ND OPINION NOTES *** Take an exposure and listen to tank. If you hear an arcing sound most likely bad tank. Rotate Gantry so tank is upside down and take exposure. Listen for arcing. Turn off lights in room and check for a flash from the tank. Remove tank, set on floor on top of the lifting bracket. Make room quiet. Rock tank front to back and side to side, listen carefully. Any sounds, indicates a failure. Check tank for oil leaks Old FRU Compatible No Comments ct/i helio rspl Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable Yes Use on Hands No Use Demand History Yes Ultimate Prime 2203602 New FRU Backward Compatible Yes Old FRU Usable No Change Type Direct ECO Number 9999
Area
CT

Full cause & step-by-step resolution