GE Healthcare · CTLightSpeed1019116896
LightSpeed 5.X (60kW)
Symptom
MOD Save or Restore functions fail. MOD Save or Restore functions fail Troubleshooting There is no media in the drive message appears in a attention box during a MOD Save or Restore of with Error: 10191 Failed save exam. posted in the Message Log. The fllowing pop-up message appears on screen. Problem is consistent If a different disk is used then the error occurs again when trying to save, restore, or label
Error message
No Archive media is attached. Request failed. Arserver No Archive media is attached. Request failed. MESSAGE from Process 9281 (ArchiveDevice.c:3192) Attach in S scsi open(..., /dev/scsi/sgc2b0t3u0) dsopen failed - errno = 2 S open(..., /dev/scsi/sgc2b0t3u0) could not open Unix d evice 16:34:34 vdi open: could not S open '/dev/scsi/sgc2b0t3u0' MESSAGE from Process 9281 (ArchiveDevice.c:4685) Arserver No Archive media is attached. Request failed. Error: 10191 Failed save exam. Mon Nov 27 03:45:02 2006 Error: 10191 Host: emc1 Process: arserver File: arscommand.c Line: 2304 Failed save exam. Paused 16896
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required