Pular para o conteúdo principal
GE Healthcare · CTLightSpeed200002632

LightSpeed RT16 & RT16 Xtra and Vision

Symptom

Date on System changed to some day in the year 2001, or earlier, after shutdown and reboot.

Error message

Error: 200002632 System Back dated. SW Options will be disabled. See Long Description for Details. Mon Dec 31 18:03:15 2001 Error: 200002632 Host: darc Process: dataacq File: swokutl.cxx Line: 784 *** Error: System Back dated. SW Options will be disabled. Please change the system date to current date or contact OLC

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2334572 The 3 VOLT BATTERY 1" DIAMETER for CPU's aka CR2032 See long description for details 2334572 The 3 Volt Battery, 1" in diameter, for CPU's also known as CR2032 This may be purchased locally at a variety of locations, including Radio Shack, Walgreens, etc. Part #: 2334572 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 3 VOLT BATTERY 1" DIAMETER FRU: Yes Status: Orderable Comments: FLS - NEWARK # 13C1019 # MANUF. CPI CANADA VC#726749 PART# 742290100 3 VOLT BATTERY 1" DIAMETER U/O LORAD M4 MICROPROCESSOR MS29328 - LMG - 2/18/02
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.