GE Healthcare · CTHiSpeed186218186328
Scanner Hardware Stopped Scan.
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 186218 (90-0704H) Rotor heater hold too long. See long description for details Error code: 0704H Rotor/Heater hold too long. Message/explanation: Rotation/Heater hold too long. Will pop up if preparation command from the system is maintained longer than 3 minutes. Potential cause: Software problem. Recommended action/Troubleshooting guide: No action Tue Aug 10 16:21:18 2010 Error: 186218 Host: GE CTRT Process: tgp out File: tgp in input.c Line: 334 XG Error phase is 3 code is 90. detail error code : 90 - 0704H Application error : Rotor/Heater hold too long. Error: 186328 13-1002-04 Gantry Req Error: ??? Servo amp Alarm is detected See the Long Description for the complete error. Error: 186328 13-1002-04 Gantry Req Error: ??? Servo Amp alarm is detected Gantry Req Error: 4 or 7 Code: 13-1002-04 Name: Servo Amp Alarm. (GP) Class: Error Description: Servo Amp alarm is detected. Potential Cause / FRU 1. Too many acceleration/deceleration in short duration. 2. Servo Amp 3. Axial Motor
Applies to
- Models
- HiSpeed
- Product
- HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
- Component
- 2197339 ( P9237HJ ) RF PLATTER POWER BRUSH BLOCK REPLACEMENT KIT see long description for details This is for the older style ETC Slip rings Please confirm which Slip Ring You have Part #: 2197339 Part Type: GP CONSUMABLES Part Weight: -226997.73 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9237HJ - RF PLATTER POWER BRUSH BLOCK FRU: Yes Status: Orderable FLS - POWER BRUSH BLOCK U/O CT HISPEED-YMS GANTRY
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.