Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000739111401125991140012600

LightSpeed 4.X - GRE

Symptom

"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors

Error message

Error: 260007391 The Data Acquisition System Has Detected An Error In The -/+12V analog DAS Power Supply See LD for more details Tue Aug 8 07:49:05 2006 Error: 260007391 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 90 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:49:05:835 Log Series: 604 The Data Acquisition System Has Detected An Error In The -12V Analog DAS Power Supply.^B Measured: 0 mV , Expected range: -11401 to -12599 mV Tue Aug 8 07:49:05 2006 Error: 260007391 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 90 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 07:49:05:735 Log Series: 604 The Data Acquisition System Has Detected An Error In The +12V Analog DAS Power Supply.^B Measured: 0 mV , Expected range: 11400 to 12600 mV

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170042P22 30A, 600V PANEL-MOUNTED FUSEHOLDER - used on GDAS See Long Discription for Details Part #: 46-170042P22 Part Type: GP CONSUMABLES Part Weight: 0.29 LBR Repairable: No Returnable: No Hazardous Material: N Description: 30A, 600V PANEL-MOUNTED FUSEHOLDER. FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.