Pular para o conteúdo principal
GE Healthcare · CTHiSpeed11074227761105710043

Cannot send a specific exam to PACS

Symptom

Cannot send a specific exam to PACS Cannot send exam to PACS with Error: 11057 Translation Failure. Genesis . Grp: 32 , Elm:82

Error message

Error: 11074 Send failed: exam/series/image (remote host) See long description for a sample error message. Mon Mar 9 10:43:44 2009 Error: 11074 Host: arhsbscti Process: dcserver File: dcscommon.c Line: 242 Send failed: 22776/2/1 (PACS)

Error: 11057 Translation Failure. Genesis . Grp: 32 , Elm:82 See LD for full error Error: 11057 Translation Failure. Genesis . Grp: 32 , Elm:82 Wed Feb 15 18:54:24 2006 Error: 11057 Host: ctge01 Process: dcserver File: dcspush.c Line: 784 Translation Failure. Genesis . Grp: 32 , Elm:82

Error: 10043 Unable to translate image to DICOM format See long description Fri Mar 9 13:52:45 2001 Error: 10043 Host: ct01 Process: arsjob File: ArchiveDevice.c Line: 7411 Unable to translate image to DICOM format, image will not be saved: /export/home1/sdc_image_pool/images/p1516/e1516/s4407158177.CTDC.290 .

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1
Area
CT

Full cause & step-by-step resolution