GE Healthcare · CTLightSpeed230017083385870381015230017078503421111522307
LightSpeed 4.X - GRE
Symptom
Fastcal is failing with beam obstructed errors. Fastcal is failing with beam obstructed errors
Error message
Error: 230017083 Possible beam obstructed - sin/cos vector magnitude = 0.xxxxxx - may include row and channel numbers See LD for complete error Mon Oct 31 07:08:03 2005 Error: 230017083 Host: cpcvct2p Process: accs File: aircal_spectral_acal_util.c Line: 445 Possible beam obstructed - sin/cos vector magnitude = 0.385870 Mon Mar 21 07:27:12 2011 Error: 230017083 Host: sj01_ct64s_6001 Process: accs File: aircal_spectral_acal_util.c Line: 449 Possible beam obstructed - row 58(26B), detector channel 375, sin/cos vector magnitude = 0.381015 Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microseconds = 108580 View scanfileto check data and header information
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5142849-4 SCAN WINDOW LIGHTSPEED SERIES H2 THRU HP RIM COVERS (replaces 2101967-4). See long description for details. Part #: 5142849-4 Part Type: GP CONSUMABLES Part Weight: 7.94 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASSEMBLY, SCAN WINDOW, LIGHTSPEED SERIES H2 THRU HP RIM COVERS, WEATHERSTRIP UPDATED MAY 2005 FRU: Yes Status: Orderable Comments: This part is replacement for 2101967-4, please discontinue use of 2101967-4 and use only 5142849-4. Added per ECR2014906, ms13452 050713
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required