GE Healthcare · CTLightSpeedVCT200900025
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Connect Pro Modality Work List will not update. Connect Pro Modality Work List will not update
Error message
Error: 200900025 Parse Failed : Invalid Worklist received from remote HIS(xxxx)(xxxx) :-xxx Note in the error example above that the error is referring to dicom tag number 0010 2154 . The numbers shown in this location of an error message like this will have the leading zeroes stripped off on tags that have leading zeroes such as the example shown. Thu Sep 28 00:01:23 2006 Error: 200900025 Host: gect1 Process: WLServer File: WLapiSystem.c Line: 191 Function: Data Management : OC Processing Patient Worklist Manager error Source File: WLS_mfm.c, Line: 1465, Status: 23 Parse Failed : Invalid Worklist received from remote HIS (10)(2154) :-113 Tue Sep 13 18:13:04 2011 Error: 200900025 Host: LS16 Process: WLServer File: WLapiSystem.c Line: 191 Function: Data Management : OC Processing Patient Worklist Manager error. Source File: WLS_mfm.c, Line: 1471, Status: 23 Parse Failed : Invalid Worklist received from remote HIS(32)(1060) :-114
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Software
- 10MW06.5 Titan for VCT (MUST HAVE SP1 or Higher) Systems running this version of software MUST also have SP1 or higher installed. Refer to FMI 25429 (SP1 release) 10MW06.5 Titan for VCT Delivered via new install or FMI 25425 Systems running this version of software MUST also have SP1 or higher installed. Refer to FMI 25429 (SP1 release)
- Area
- CT