GE Healthcare · CTBrightSpeedVCT2600074285035926000741620053260007774260007429
BrightSpeed HPower, Delight, Velocity or MERI HPower
Symptom
Customer sees "Detector Temperature Is High. Image quality may be affected". Customer sees Detector Temperature Is High. Image quality may be affected
Error message
Error: 260007428 WARNING: DAS Temperature is nearing the Upper Limit. See LD for more details Mon Aug 28 06:39:06 2006 Error: 260007428 Host: orp Process: orp_subsys File: ctemptst.c Line: 600 Mon Aug 28 06:39:04 2006 Host : Das Ermes # : 260007428 Exception Class : Soft Severity : Sec/Soft File : ctemptst.c Line# : 600 Function : No System Function Reported Scan Type : None/Unknown Scan: 50359/1/0 WARNING: DAS Temperature is nearing the Upper Limit. Highest temperature was seen is on Converter Board 23. Measured: 63.0 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. See Following Error Message for Converter Card List Error: 260007416 Detector Temperature Is High. Image quality may be affected See Long Description Tue Sep 6 09:07:45 2005 Error: 260007416 Host: orp Process: orp_subsys File: dtemptst.c Line: 364 Tue Sep 6 09:07:44 2005 Host : Das Ermes # : 260007416 Exception Class : Soft Severity : Sec/Soft File : dtemptst.c Line# : 364 Function : No System Function Reported Scan Type : Scout Scan: 372/1/1 Detector Temperature Is High. Image quality may be affected. Current Reading:40.41 Deg C A/D Counts:20053 Error: 260007774 List of Converter Cards From the Previous Error Me
Applies to
- Models
- BrightSpeed, VCT
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5340738 (replaces 2359114) HPOWER GANTRY FAN AND HARNESS used on 5.x gantry See Long Description for details This fan for BrightSpeed, HP100, HP60 and RT Systems. Part #: 5340738 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GANTRY FAN AND HARNESS FRU: Yes Status: Orderable Comments: Per ECR 2091297, 212008898, 10/26/09. Part #: 2359114 <-- All GPRS stock is gone and this part is now obsolete Part Type: GP CONSUMABLES Part Weight: 8.82 LBR Repairable: No Returnable: No Hazardous Material: N Description: HPOWER GANTRY FAN AND HARNESS FRU: Yes Status: Orderable Comments: FLS - HPOWER GANTRY FAN AND HARNESS.ADDED PER THE HPOWER RSPL. EXPECTED M3 FOR THIS PROGRAM: FW19, 2003. MS20544 2/28/03.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required