GE Healthcare · CTLightSpeed2300110755000082000102000

LightSpeed 4.X - GOC2

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230011075 Unable to open a scan file on the scan data disks. Read of scan file header failed See the LD for error codes Error: 230011075 Unable to open a scan file on the scan data disks. Read of scan file header failed in read Thu Dec 29 07:56:06 2005 Error: 230011075 Host: srct Process: scanmgr File: sdio_xfs_io_ops.c Line: 662 Function: Data Acquisition : SBC Processing Scan Type: Axial Scan: 50000/8/13 Scan Seq Id: 82000 Unable to open a scan file on the scan data disks. Read of scan file header failed in read() call. Scan: CT01/50000/8/13 Bytes to read: 198h Starting offset: 0h Number read: 0h OS error : 0 --- Success Or Fri Oct 6 08:27:13 2006 Error: 230011075 Host: ct01 Process: scanmgr File: sdio_xfs_io_ops.c Line: 662 Function: Data Acquisition : SBC Processing Scan Type: Helical Scan: 5144/3/1 Scan Seq Id: 102000 Unable to open a scan file on the scan data disks. Read of scan file header failed in read() call. Scan: CT01/5144/3/1 Bytes to read: 198h Starting offset: 0h Number read: ffffffffh OS error : 11 --- Resource temporarily unavailable

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.