GE Healthcare · CTLightSpeedVCT260118220
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118220 (60-0325) DSP overload See Long Description Sun May 30 15:50:49 2010 Error: 260118220 Host: orp Process: orp_subsys Host : Jedi Ermes # : 260118220 Exception Class : Abort Severity : Primary Scan Type : Helical Scan: 9451/2/2 . (60-0325)DSP overload Code 60 Exposure Errors Error code: 60-0325H Message/explanation: DSP overload Default in kV regulation calculation Potential cause: Main supply Inverter current measurement PPC kV control board Inverter HV tank defective Check main DCBUS voltage on inverter Recommended action/ Troubleshooting guide: Check if transformer currents measurement is correctly plugged inside the inverter (J1 in GateCmd board or J19 in CT IF board). Take some exposures. If error persists, investigate those 2 root causes: 1. Inverter / kV Ctrl: run Inverter in Short Circuit diagnostic. If test fails replace PPC kV control board, if this test fails again after replacement of PPC kV ctrl Board, replace the Inverter. 2. HV Tank: run No Load HV diag without removing HV cables. If test fails, replace HV tank Note: When filament cals are run on tube with open filament errors, the error is generated with DSP overlaod.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5136161-4 PPC KV Control replaces (5136161, 5136161-2, 5117564-10) used by all 5.X, Brightspeed and VCT gantries See Long Description for Details PPC KV Control Bd. 5136161-4 replaces (5136161, 5136161-2, 5117564-10) this is the orderable Programmed KV Control Board for HP60, LSPro 80KW , 100KW and VCT 7.X, RT16, BrightSpeed Power and Select Systems (Velocity, Delight, Meri) Part #: 5136161-4 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED kV CONTROL P6.33BF6 AS FRU FRU: Yes Status: Orderable Current: 5136161-4 Harvested: Not Available Exchange: Not Available Comments: Jedi KV Control Board used on Definium 8000, Revolution XR/d-2X (Lightning), Definium 6000, and Precision 500D. Also may be used on all Innova systems Bolus/Unity/SPAD/Harmony. P.Liebau 1/2010 Part #: 5136161-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED kV CONTROL 6.30BFA AS FRU FRU: Yes Status: Orderable Comments: In order to avoid Jedi software download issues on Definium 8000, Revolution XR/d-2X (Lighting), Definium 6000, and Precision 500D you may order 5136161-4, KV Control Board developed for these systems. Part #: 5136161 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED KV CTRL P6.30BFA as FRU FRU: Yes Status: Orderable Comments: (2/21/06 212001261) The 5136161 KV board replacement is the preferred part used on all HP60 hp 80 LS-RT and VCT products.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.