Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2005020032100004413863552210000434200109012

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Unable to scan axial or helical; system cannot allocate image space as one chunk. Unable to scan axial or helical; system cannot allocate image space as one chunk

Error message

Error: 200502003, Cannot allocate image space for XXX images, allocated XX images. See LD for complete error Mon Feb 13 16:36:28 2006 Error: 200502003 Host: ls16dn Process: retroRecon File: srx_sma_manage_image_space.c Line: 690 Function: No system function reported Cannot allocate image space for 148 images, allocated 7 images.

Error: 210000441 Image space allocation failed. See LD for complete error Mon Mar 26 12:51:14 2007 Error: 210000441 Host: roct Process: dms File: Initialize.cxx Line: 856 Image space allocation failed. size 3863552 Error code: -1

Error: 2106 Image Storage Full See Long Description for complete error Wed Aug 30 16:07:53 2006 Error: 2106 Host: IRIS Process: imserver File: imsprim.c Line: 742 Image Storage Full

Error: 210000434 IMAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data See Long Description for complete error Fri Apr 21 07:41:23 2006 Error: 210000434 Host: lightspeed Process: dms File: SaveContext.cxx Line: 184 I MAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data.

Error: 200109012 Image Allocation Failed. Cannot build scan request. See Long Description for complete error Wed Mar 2 18:39:33 2005 Error: 2001

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required