GE Healthcare · CTLightSpeed26011820924097
LightSpeed 5.X (60kW)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118209 (60-0311) ILP current not OK See LD for error explanation Sun Nov 30 10:50:13 2008 Error: 260118209 Host: orp Process: orp_subsys File: gen_exec.c Line: 2446 Sun Nov 30 10:50:13 2008 Host : Jedi Ermes # : 260118209 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2446 Function : No System Function Reported Scan Type : Helical Scan: 24097/2/1 (60-0311)ILP current not OK. The current in the parallel resonant circuit of the inverter did not rise at the beginning of the exposure. Z6,Z7= 0h,0h Message/explanation: ILP current not OK. The current in the parallel resonant circuit of the inverter did not rise at the beginning of the exposure.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5136161-4 PPC KV Control replaces (5136161, 5136161-2, 5117564-10) used by all 5.X, Brightspeed and VCT gantries See Long Description for Details PPC KV Control Bd. 5136161-4 replaces (5136161, 5136161-2, 5117564-10) this is the orderable Programmed KV Control Board for HP60, LSPro 80KW , 100KW and VCT 7.X, RT16, BrightSpeed Power and Select Systems (Velocity, Delight, Meri) Part #: 5136161-4 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED kV CONTROL P6.33BF6 AS FRU FRU: Yes Status: Orderable Current: 5136161-4 Harvested: Not Available Exchange: Not Available Comments: Jedi KV Control Board used on Definium 8000, Revolution XR/d-2X (Lightning), Definium 6000, and Precision 500D. Also may be used on all Innova systems Bolus/Unity/SPAD/Harmony. P.Liebau 1/2010 Part #: 5136161-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED kV CONTROL 6.30BFA AS FRU FRU: Yes Status: Orderable Comments: In order to avoid Jedi software download issues on Definium 8000, Revolution XR/d-2X (Lighting), Definium 6000, and Precision 500D you may order 5136161-4, KV Control Board developed for these systems. Part #: 5136161 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED KV CTRL P6.30BFA as FRU FRU: Yes Status: Orderable Comments: (2/21/06 212001261) The 5136161 KV board replacement is the preferred part used on all HP60 hp 80 LS-RT and VCT products.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required