GE Healthcare · CTLightSpeedVCT29000900931809290010005
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Recon self test failure on boot up; scanning would be allowed but no reconstruction of images. Recon self test failure on boot up; scanning would be allowed but no reconstruction of images
Error message
Do not Use - use one of the other Error: 290009009 Image failed due to many reasons See Long Description for complete error Tue Apr 19 12:44:49 2005 Error: 290009009 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 3968 Function: Image Generation : SBC Processing Job Type: RECON_JOB_PROS_AXIAL_RECON Scan: 4354/2/3 Image: 4354/2/6 Image failed due to: IG error: PROCESSING_ERROR. Number of retry so far: 1. Tue Nov 7 11:38:44 2006 Error: 290009009 Host: darc Process: recon control File: ImageSetRequest.cxx Line: 5348 Function: Image Generation : SBC Processing Job Type: RECON JOB PROS AXIAL RECON Scan: 7695/4/2 Image: 7695/4/2 Image failed due to: IG Lost Connection. Number of retry so far: 0. IG node: 3. Tue Jun 26 12:44:49 2005 Error: 290009009 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 3968 Function: Image Generation : SBC Processing Job Type: RECON_JOB_PROS_AXIAL_RECON Scan: 4354/2/3 Image: 4354/2/6 Image failed due to: image installed with status -1. Number of retry so far: 0. IG node: 0. Fri Jan 18 06:03:34 2008 Error:290009009 Host: darc Process: recon_ control File: ImageSetRequest.cxx Line: 6076 Function: Image Generation : SBC Processi
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Software
- 09MW08.11 for GOC6 LightSpeed VCT 7.2 (64 Slice Only) . This software is now obsolete, this site should be upgraded to 11HW12.5 software released via FMI 25427 09MW08.11 for GOC6 LightSpeed VCT 7.2 (64 Slice Only) Vision Refer to SN09051913a LightSpeed 7.2 ( 09MW08.11 ) Software Release supports GOC6 consoles used on VCT 64, VCT XT, VCT XTe and VCT Std. - LightSpeed 7.2 (09MW08.11 )Software for GOC6 Collector 5311595 Rev 3 Includes OS and Application CD's - Operating System Linux 5.3.8 Software CD 5323584 Rev 1 - Application LightSpeed VCT GOC6 DVD 5336533 Rev 3 - Class C LightSpeed 7.2 5336535 Rev 2 - Class M LightSpeed 7.2 5336536 Rev 2 An On-site/Remote identification aid . (Example: check_config) Result from check_config or swhwinfo 09MW08.11.V40_H_V64_G_GTM Please create an ITRAK report for any software related issues
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.