GE Healthcare · MROptima22238812000012815350602269307
Application Software does not complete the boot process with TPS Download failure.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2223881 Host to AGP board ethernet communication timed out during TPS Reset. Aug 11 Wed 12:55:17 2223881 NSP ../TardisReset.cc 1898 Exception Class: Unknown Function: Host: NSP TPS Reset Communication Failure: Host to AGP board ethernet communication timed out during TPS Reset.-> Value 0xAAAA not found at address 0x800000A8 after 20000 msec.-> 1281535060 0 1 2269307 Communication problem between the AGP and SCP boards. 2269307 diagCoord pearlMap.c 2171 Task ID: 0x0 Time: 0 Object: SCP Exception Class: Unknown Function: Software/Firmware/Config: N/A Communication problem between the AGP and SCP boards.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- 21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- SRF/TRF Board was hanging the MGD bus.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- IApplication Software does not complete the boot process with TPS Download failure Due to Idle Overnight With Power On.