Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23000006450003164000

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

"Scan hardware Stopped Scan" with Taxi data violation errors. "Scan hardware Stopped Scan" with Taxi data violation errors

Error message

Error: 230000064 TAXI data violation detected by forward error correction See the long description for the full error code Sat Apr 17 20:26:43 2004 Error: 230000064 Host: darc Process: dataacq File: dip_thread.cxx Line: 411 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Axial Scan: 50003/5/1 Scan Seq Id: 164000 TAXI data violation detected by forward error correction. Interrupt bit: 4 Violation count: 1 Scan aborted.

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2205268 15V 1.7A power supply - Used In 1.X, 2.X, 3.X, 4.X, and 5.X gantrys 2205268 15V 1.7A power supply Used 1.X, 2.X, 3.X, 4.X, 5.X gantrys Part #: 2205268 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: POWER SUPPLY 15V DC 1.7 AMPS FRU: Yes Status: Orderable Comments: FLS - POWER SUPPLY 15V DC 1.7 AMPS U/O HELIOS GANTRY MS08992 10/5/98 USED ON HPOWER PROGRAM, PER RSPL. MS20544 ON 2/12/03. PART TRANSFERRED TO SSP 04/21/00 MS24085 PER KINH LEE CT MFG. THIS POWER SUPPLY IS FOR BOTH THE TRANSMITTER, (MOUNTED TO SLIPRING), AND ALSO RF RECEIVER (MOUNTED TO STATIONARY GANTRY). MS10604 001218 NOTE: F/O CABLE IS 2261709
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.