GE Healthcare · CTLightSpeed2600079315003026000740250444
LightSpeed RT16 & RT16 Xtra and Vision
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260007931 DAS Converter Board(s) failed to write the ASIC Trim Table see long description Fri Jul 13 07:08:44 2007 Error: 260007931 Host: orp Process: orp_subsysflt.c Line: 82 Host : Das Ermes # : 260007931 Exception Class : Secondary Severity : Sec/Soft File : brdflt.c Line# : 82 Function : No System Function RReported Scan Type : Axial Scan: 50030/1/3 DAS Converter Board(s) failed to write the ASIC Trim Table See Following Error Message for Converter Card List Error: 260007402 Converter Board(s) Asserted a Board FAULT. See Long Description for complete error Tue Dec 28 09:41:31 2004 Error: 260007402 Host: orp Process: orp_subsys File: cnvmgr.c Line: 605 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260007402 Exception Class : Abort Severity : Secondary File : cnvmgr.c Line# : 605 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 Converter Board(s) Asserted a Board FAULT.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2404393 GDAS-16 CONVERTOR CARD, replaces 2364354. See Long Description for details. 2404393 GDAS-16 CONVERTER CARD, replaces 2364354 Part #: 2404393 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: 16 SLICE GDAS CONVERTER CARD FRU, A CONVERTER CARD IN AN ESD BAG IN A BOX WITH A PART NUMBER LABEL FRU: Yes Status: Orderable Comments: FLS - 16 SLICE GDAS CONVERTER CARD USED ON: LIGHTSPEED PRO 16 FRU GDAS ADDED PER GDAS RSPL MS13452 040315 RISK: SITE DOWN FREQUENCY: HIGH NOTE: THERE ARE 96 OF THESE CARDS PER SYSTEM, BUT SHOULD NOT NEED TO REPLACE ALL AT ONETIME, MS13452 THIS PART CONTAINS ONE GDAS CONVERTER CARD 2364354 WITH FRU PACKAGING, SUPPLIED BY GEMS IT. MS13452 PDR 8471.80.9000 IM05 UPDATE AWAITING MM (212001261 4/1/06) For 4/8 slice G Das which uses a thick connector must order 2364354-2
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.