Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18333819507183270183262183337

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183338 The maximum allowable HVDC Rail voltage drop has been exceeded. See LD for complete error Tue Jul 25 09:43:35 2006 Error: 183338 Host: OBC Process: Generator File: gen_power.c 1.13 Line: 258 Function: SYSTEM MONITORING : HVDC Monitor Scan: 19507/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 09:43:35:145 Log Series: 528 The maximum allowable HVDC Rail voltage drop has been exceeded. No load HVDC value: 663V 1st HVDC sample: 27V HVDC 15mS later: 27V Rotor LOV bit value: 1 From Feedback If error indicates approximately half of proper HVDC supply, this indicates strong chance of blown 700V fuse F1 or F2 on HEMRC I/F board in HEMRC

Error: 183270 Backup contactor sense error. See LD for more details Error: 183270 Backup contactor sense error. HVDC Bus was up, but the I/O board indicated that the contactor was de-energized. BCSTAT = 0

Error: 183262 Backup contactor opened without firmware command See the long description for the full error Mon May 17 12:20:18 1999 Error: 183262 Suite: CT01 Host: OBC Process: Generator File: ctvrc_manager.c 1.8 Method: No Method Line:823 Function: XRAY GENERATION : 550 VDC Backup Contactor Intlk Scan: 7073/3/15 Type: Axial Exceptio

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2285089 SLIPRING ASSY, HSDCD 400MG PLATTER See long description for details Part #: 2285089 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: SLIPRING ASSY, HSDCD 400MG PLATTER FRU: Yes Status: Orderable Comments: FLS - SLIPRING ASM, HSDCD 400M. USED ON WARP 3.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues