GE Healthcare · CTLightSpeedVCT260118192
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan with open filament error in log. Scanner Hardware Stopped Scan with open filament error
Error message
Error: 260118192 (50-0211)Jedi detected open filament circuit. See Long description for full error Host: Jedi Ermes #: 260118192 Exception Class: Abort Severity: Primary File: gen_exec.c Line#: 2148 Function: No System Function Reported Scan Type: Helical Scan: 542/2/1 (50-0211) Jedi detected open filament circuit. Verify tube, HV cable, HV tank filament transformer, and Jedi interconnect cables are properly installed. RMS filament current measurement (every 0.5 msec) on heater board too high. Z6,Z7= 97h,0h
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2361419 Programmed Heater Board (eqvt 2409049) inside Jedi Auxiliary Module 2324118 See LD for Part Details - please read the note at the bottom. Part number for Heater (Filament) Board below: Part #: 2361419 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED HEATER 6P04 FRU: Yes Status: Orderable Part #: 2409049 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: FRU-PROGRAMMED HEATER V6.04 FRU: Yes Status: Orderable Parts Offered as: Current: 2409049 Harvested: Not Available Exchange: Not Available Comments: added per to ECR2088514 by sso#305010201 in Jan 15, 2010 Note: * Order 2409049 to replace 2361419.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.