GE Healthcare · CTBrightSpeed260121013266942601210212050020316
BrightSpeed HPower, Delight, Velocity or MERI HPower
Symptom
Gantry angle number stuck at (some value) Gantry angle number stuck at S26.5
Error message
Error: 260121013 Gantry velocity less than the minimum allowed velocity. See Long Description for more details Wed Mar 29 11:28:24 2006 Error: 260121013 Host : Table Gantry Processor Ermes # : 260121013 Exception Class : Secondary Severity : Pri/Soft File : tilt mgr.c Line# : 1171 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Gantry velocity less than the minimum allowed velocity. Gantry velocity:-2 mdeg/sec Tilt position:-26694 mdeg Minimum allowable velocity:300 mdeg/sec Error: 260121021 Tilt Stopping Error. Tilt did not reach target angle See LD for complete error Fri Dec 21 16:47:02 2007 Error: 260121021 Host: tgp Process: tgp_subsys File: tilt_monitor.c Line: 299 Fri Dec 21 16:47:01 2007 Host : Table Gantry Processor Ermes # : 260121021 Exception Class : Secondary Severity : Pri/Soft File : tilt_monitor.c Line# : 299 Function : No System Function Reported Scan Type : Helical Scan: 7649/2/1 Tilt Stopping Error. Tilt did not reach target angle -20500 mdeg. Tilt current angle: -20316 mdeg. Tilt angle tolerence: 250 mdeg. Timeout: 1000 msec.
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2348298 GANTRY TILT TIMING BELT, 2.03 PITCH X 6 WIDE -used on LightSpeed 5.X and 7.X see LD for details P/N 2348298 GANTRY TILT TIMING BELT, 2.03 PITCH X 6 WIDE -used on LightSpeed 5.X and 7.X picture below Part #: 2348298 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TIMING BELT, 2.03 PITCH X 6 WIDE FRU: Yes Status: Orderable Parts Offered as: Current: 2348298 Harvested: Not Available Exchange: Not Available Comments: FLS - TIMING BELT. ADDED PER THE HPOWER RSPL. EXPECTED M3 FOR THIS PROGRAM: FW19. MS20544 ON 3/24/03 5122696 Tilt Pot with Harness See long description for details Part #: 5122696 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: Tilt Pot with Harness FRU: Yes Status: Orderable Comments: Added per EMC2 RSPL.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required