Symptom

ECG Trace fails to display on the console. See Long Description for more details. ECG Trace intermittently fails to display on the console, requiring a software reboot to recover. Problem seen after upgrading to 06MW03.4 software. The ECG trace works the first time after a reboot but after any interruption, the trace cannot be restored using the Gating BPM button.

Error message

Error: 200501547 A Communication Error has occurred in uifIpcSendRecvWaveFormViewerShow with seq num = 33 and status = 200001821 See long description for a sample error message. Wed Feb 17 10:36:37 2010 Error: 200501547 Host: kghcate vct Process: scanRx File: uifIpcMgrScanRx.c Line: 12671 Function: Data Acquisition : OC Processing A Communication Error has occurred in uifIpcSendRecvWaveFormViewerShow with seq num = 33 and status = 200001821

Error: 200501547 A Communication Error has occurred in rxif show external waveform viewer() with seq num = 200001821 and status = -1078817464 See long description for a sample error message. Wed Feb 17 10:36:37 2010 Error: 200501547 Host: kghcate vct Process: scanRx File: SubSystemIf.c Line: 4639 Function: Data Acquisition : OC Processing A Communication Error has occurred in rxif show external waveform viewer() with seq num = 200001821 and status = -1078817464

Error: 200501532 uif_scan_rx failed to receive message from service ekg_broker_svc See long description Fri Dec 11 15:27:00 2009 Error: 200501532 Host: bay43 Process: scanRx File: uifIpcCommonFunctions.c Line: 2508 Function: No system function reported Service uif_scan_rx failed to rece

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.