Scanner Hardware Stopped Scan.

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260105022 (25-0555) JEDI-SC:The main power of SC Drive(48V) or SC Supply (96V) is too low. Others: LVPS TRI: detected +160 V too low. See Long Description for full error Mon Aug 16 21:39:25 2010 Error: 260105022 Host: orp Process: orp_subsys File: gen_exec.cxx Line: 2792 Mon Aug 16 21:39:24 2010 Host : Jedi Ermes # : 260105022 Exception Class : Secondary Severity : Sec/Soft File : gen_exec.cxx Line# : 2792 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 . (25-0555) JEDI-SC:The main power of SC Drive(48V) or SC Supply (96V) is too low. Others: LVPS TRI: detected +160 V too low.

Applies to

Models
Discovery
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5197802-2 Cherokee 5kW 48vdc power supply with fuse board. NOT USED ON VCT. Used on on a HD-CT750 system gantry rotating frame. 5197802-2 is used on a HD-CT750 system, Do not use on a VCT system Part #: 5197802-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCTHD 48V PS with FRU packaging FRU: Yes Status: Orderable Comments: **FE ASSIST** 5197802-2 is used on a HD-CT750 system, Do not use on a VCT system. (212001261 9/30/2010) .Per HDVCT RSPL, 212008898, 11/6/08
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.